AlarmMEC Board — Power Supply Rails

HC1100324V / 12V / 5V voltages not OK on the MEC board

Agfa CT error code, reported by the MEC Board.

What HC11003 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11003 indicates: 24V / 12V / 5V voltages not OK on the MEC board. The code originates in the MEC Board — Power Supply Rails section of the MEC Board. Severity level alarm (Alarm).

Reported action: Check the connection first: connector P24 on the MEC board, connector P3 on the power supply, and the wiring from PSU-P3 to MEC-P24. If the connection is good, the fault is either on the power supply or in the voltage-measuring function of the MEC board — replace the failing one..

Raised on the Drystar 5500/5503 when the 24V, 12V and 5V voltages are not OK on the MEC board. The printer shows "Call service". Catalogued in the appendix next to the MEC board's other power fault codes, HC11001 and HC11002.

Full troubleshooting for HC11003