AlarmMEC Board — Power Supply Rails

HC1100124V actuator power not available on the MEC board

Agfa CT error code, reported by the MEC Board.

What HC11001 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11001 indicates: 24V actuator power not available on the MEC board. The code originates in the MEC Board — Power Supply Rails section of the MEC Board. Severity level alarm (Alarm).

Reported action: Check the connection first: connector P25 on the MEC board, connector P4 on the power supply, and the wiring from PSU-P4 to MEC-P25. If the connection is good, the fault is either on the power supply or on the MEC board — replace the failing one..

Raised on the Drystar 5500/5503 when the 24V power for the actuators is not available on the MEC board. The printer shows "Call service". Catalogued in the appendix next to the MEC board's other power fault codes, HC11002 and HC11003.