HFO Table Will Not Lower.
Symptom
HFO Table Will Not Lower. See Long Description There are two different optic sensors tied to the same set block at the rear of the table. If the sensors are not aligned with the each other, they will trip at different points. Table Home allows the table to be lowered and SRI Home tells the cradle to stop moving out (it has reached home, or is near the end of travel out of the bore). It is essential that the system sees the Table Home sometime before or at the same time as the SRI Home, but not before it is safe to lower the table (i.e. not before cradle is out sufficiently, such that when table is lowered cradle won't catch on bridge). If it never sees the Table Home, the cradle will not be allowed to lower.
Error message
2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Multicoil Bias Fault errors on multiple coils.
- Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
- The system is unable to scan with any surface coil, get dynamic disable faults.
- Auto Prescan Fails in Body Transmit Mode With TG Reached the High Limit due to HFO Bottom RF Body Coil Assembly.
- Getting intermittent SCIM and Keyboard Errors.