Pular para o conteúdo principal
GE Healthcare · MR2224666

HFO Table Will Not Lower.

Symptom

HFO Table Will Not Lower. See Long Description There are two different optic sensors tied to the same set block at the rear of the table. If the sensors are not aligned with the each other, they will trip at different points. Table Home allows the table to be lowered and SRI Home tells the cradle to stop moving out (it has reached home, or is near the end of travel out of the bore). It is essential that the system sees the Table Home sometime before or at the same time as the SRI Home, but not before it is safe to lower the table (i.e. not before cradle is out sufficiently, such that when table is lowered cradle won't catch on bridge). If it never sees the Table Home, the cradle will not be allowed to lower.

Error message

2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues