The system is unable to scan with any surface coil, get dynamic disable faults.

Symptom

The system is unable to scan with any surface coil, get dynamic disable faults. Head and Body coil scans do not produce this error.

Error message

2225541 Dynamic disable fault, Rear of Coil open. Dynamic Disable bias fault on back drivers. Wed Dec 1 09:19:34 2004 Error: 2225541 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6160 Dynamic disable fault, Rear of Coil open. Dynamic Disable bias fault on back drivers. Possibly a disconnected cable between the driver board and the dynamic disable switch boxes, or an open diode in dynamic disable switch boxes on BACK of BRM.

2225539 Dynamic disable fault, Front of Coil open. Dynamic Disable bias fault on front drivers. Wed Dec 1 09:09:29 2004 Error: 2225539 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6140 Dynamic disable fault, Front of Coil open. Dynamic Disable bias fault on front drivers. Possibly a disconnected cable between the driver board and the dynamic disable switch boxes, or an open diode in dynamic disable switch boxes on FRONT of BRM.

2224666 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan. Wed Dec 1 09:09:30 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan.

2224583 Auto Prescan Failed

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2111443 1 KV Dynamic Disable Filter Module
Area
MR