GE Healthcare · MRExcite224858512090559922248590222466622485841207812070
Getting intermittent SCIM and Keyboard Errors.
Symptom
Getting intermittent SCIM and Keyboard Errors. Getting intermittent errors SCIM and Keyboard errors filling the error log. Also, intermittent Keyboard initialization failures.
Error message
2248585 Repeated SCIM Communication Errors. Thu Apr 24 16:52:21 2008 2248585 2 NSP SCP:KeyboardLinkHWO 1913 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications Repeated SCIM communication errors recorded. The SCP/SCIM cable may be bad or the SCIM may be bad. Please run the keyboard (SCIM) diags. Check cable connection to Workspace Interface Module and to System Cabinet.-> EN 450 SR 217 1209055992 2248590 Keyboard Communication Error. Error: 2248590 Host: mrs1_oc0 Process: NSP File: SCP:KeyboardLinkHWO Line: 1701 Keyboard Communication Error. The SCP received a packet indicating that keyboard has reset. Check cable connection between keyboard and Workspace Interface Modul e and cable connection between keyboard and System Cabinet. 2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure. 2224666 Auto Prescan failed, Please Try Again. Error: 2224666 Host: signa1 Process: psc File: PscMainControl.cc Line: 1351 Auto Prescan failed, Please Try Again. 2248584 SCIM Communication Error. Thu Apr 10 07:21:09 2008 2248584 2 NSP SCP:Key
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 5120763 Run Number 1255, NOT A FRU System interconnect cable between System Cabinet (MR2-A11-J26) and Operator Workspace (OW1-A21-J18). Part is NOT a FRU and is NOT Orderable. Cable needs to be acquired directly from MR. System interconnect cable between System Cabinet (MR2-A11-J26) and Operator Workspace (OW1-A21-J18). Part is NOT a FRU and is NOT Orderable. Cable needs to be acquired directly from MR.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.