Pular para o conteúdo principal
GE Healthcare · MRExcite224858512090559922248590222466622485841207812070

Getting intermittent SCIM and Keyboard Errors.

Symptom

Getting intermittent SCIM and Keyboard Errors. Getting intermittent errors SCIM and Keyboard errors filling the error log. Also, intermittent Keyboard initialization failures.

Error message

2248585 Repeated SCIM Communication Errors. Thu Apr 24 16:52:21 2008 2248585 2 NSP SCP:KeyboardLinkHWO 1913 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications Repeated SCIM communication errors recorded. The SCP/SCIM cable may be bad or the SCIM may be bad. Please run the keyboard (SCIM) diags. Check cable connection to Workspace Interface Module and to System Cabinet.-> EN 450 SR 217 1209055992

2248590 Keyboard Communication Error. Error: 2248590 Host: mrs1_oc0 Process: NSP File: SCP:KeyboardLinkHWO Line: 1701 Keyboard Communication Error. The SCP received a packet indicating that keyboard has reset. Check cable connection between keyboard and Workspace Interface Modul e and cable connection between keyboard and System Cabinet.

2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure.

2224666 Auto Prescan failed, Please Try Again. Error: 2224666 Host: signa1 Process: psc File: PscMainControl.cc Line: 1351 Auto Prescan failed, Please Try Again.

2248584 SCIM Communication Error. Thu Apr 10 07:21:09 2008 2248584 2 NSP SCP:Key

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
5120763 Run Number 1255, NOT A FRU System interconnect cable between System Cabinet (MR2-A11-J26) and Operator Workspace (OW1-A21-J18). Part is NOT a FRU and is NOT Orderable. Cable needs to be acquired directly from MR. System interconnect cable between System Cabinet (MR2-A11-J26) and Operator Workspace (OW1-A21-J18). Part is NOT a FRU and is NOT Orderable. Cable needs to be acquired directly from MR.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.