GE Healthcare · MRExcite22246662254235
Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
Symptom
Auto prescan failures with RF Amp setup incomplete message
Error message
2224666 Auto Prescan failed, RF Amp setup incomplete 2224666 Host: SE3TMR145 Process: psc File: PscMainControl.cc Line: 1354 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, RF Amp setup incomplete 2254235 3T RF Amplifier PA Bias Voltage Low (fault 14). The voltage required to bias the PA tube is too low. 2254235 Host: SE3TMR145 Process: NSP File: SCP:../rf nb contro Line: 1491 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: 8HRBRAIN Scan: 9789/11 3T RF Amplifier PA Bias Voltage Low (fault 14). The voltage required to bias the PA tube is too low.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- SecondOpinion MKS 3T 35kW RF Amp 5104896 See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.