GE Healthcare · MR2225691222570322256972224790
Clarify 028125 Profile 2.
Symptom
No Display on the SCA Board and the 5VDC Indicator LED Not Lit. Error log reading 2224790, 2225691, 2225697, 2225703. All these refer to MDS link failures to the gradient system. These were coupled with the display on the SCA being off and the 5V indicator light out.
Error message
2225691 SPI MDS Link Failure. The SPI failed to receive an MDS transmission echo. 2225691 SPI MDS Link Failure The SPI failed to receive an MDS transmission echo. 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics 2225697 SPI MDS Link Failure 2224790
Applies to
- Product
- Profile 2 SPARC20 with AN8290Y gradients
- Software
- Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
- Area
- MR