GE Healthcare · MRSignaOvationExcite222569122253342225337224458522257042225697
CWA ID108152. Fails TPS Reset when booting up to Applications or when Resetting TPS.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2225691 MDS link failure. The IPG failed to receive an MDS transmission echo 2225691 MDS link failure. The IPG failed to receive an MDS transmission echo. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0006 Transmitted byte: 0x11 2225334 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Error: 2225334 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri resp timeout Line: 12 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Data packets waiting for a response: 0x4000 2225337 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Error: 2225337 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri watchdog fail Line: 13 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Watchdog packets waiting for a response: 0x0001 2244585 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready 2244585 "The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. If ENABLE test point on the front SGA-PS J2 connector is HIGH, replace the SGA-PS. Else, check the GP to SGA-PS c
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.