CWA ID115238. Scan Aborts with MDS Link Error.

Symptom

Scan Aborts with MDS Link Error

Error message

2225693 MDS link failure. An MDS device failed to respond to IPG. Wed Oct 29 00:14:08 1997 2225693 1 CMR1 OC0 HPC_PORT_ID PMH.m 1.29 spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0252 Transmitted header: 0x10 Transmitted opcode: 0x02

2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.

2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and RF Cabinets 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics.

Applies to

Models
Signa, Ovation, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.