GE Healthcare · MR2240188

Cannot scan, getting Preamp Bias faults on all Coils.

Symptom

Cannot scan, getting Preamp Bias faults on all Coils.

Error message

2240188 Pre amplifier bias error was occurred on 0 degree port. Host: oc001 Process: NSP File: HPC PORT ID:spi srp Line: 1142 Pre amplifier bias error was occurred on 0 degree port. It may be neccessory to change the pre-amplifier box.

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Component
2195463 Preamplifier box with Phase shifters Preamp must be tuned after replacement. See Rev.13 GE Yokogawa Service CD 2220946. Select Profile3/Profile4>Replacement>Table>Preamp Box Replacement>Step 4, click on "How" button on middle of page.
Area
MR

Full cause & step-by-step resolution