cannot scan with knee coil...getting preamp bias fault

Symptom

Cannot scan, getting Preamp Bias fault on one Coil.. Was able to scan with other coils, so issue isolated to a defective coil. Tried hard rebooting entire system and errors repeated with defective coil.

Error message

2240188 Pre amplifier bias error was occurred on 0 degree port. Host: oc001 Process: NSP File: HPC PORT ID:spi srp Line: 1142 Pre amplifier bias error was occurred on 0 degree port. It may be neccessory to change the pre-amplifier box.

2240187 Undefined error code received from pre-amplifier. Wed Jun 29 10:40:41 2005 Error: 2240187 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_srp Line: 1142 Undefined error code received from pre-amplifier.

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required