Pular para o conteúdo principal
GE Healthcare · MR222466622247092224580

Banana plug on the Head T/R switch has bent/broken fingers causing coil to disconnect.

Symptom

auto prescan fails with signal too small on head coil auto prescan fails with signal too small on head coil.

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver (X + 1) while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur.

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Product
3T Eclipse For E2 software
Software
VH3 3T VH3
Component
2362992 3T CORONA HEAD ASSEMBLY 2362992 3T CORONA HEAD ASSEMBLY. 3T Head T/R Switch.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.