GE Healthcare · MR22482172847174
Receiver calibration error(s).
Symptom
Warning Receiver 0 is out of spec, Image quality may be compromised message Warning Receiver 0 is out of spec, Image quality may be compromised message.
Error message
2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR5-HFO5 Open Speed with Excite Product
- Component
- 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- RRF Calibration Error Due to a Bad Connection From RRF MUX to RRF Exciter.
- RRF Calibration Error in RRF Data Path Diagnostics due to DRF board.
- Will not scan with error 2248217 Receiver calibration error.
- Receiver Cal and Low or no Signal Strength Seen From Exciter errors due to 5 V Power Supply in RRF.
- Will not scan with error 2248217 Receiver calibration error Due to Receiver Board.
- Receiver Cal Error due to DCERD2 4 channel receiver board.