GE Healthcare · MRSigna2247384185712247386224816322482172847174

Will not scan with error 2248217 Receiver calibration error Due to Receiver Board.

Symptom

Will not scan with error 2248217 Receiver calibration error Fri Jul 18 05:59:36 2003 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APSReceiver calibration error(s). Receiver DB out of spec.R1 values that failed:Rx1: 12 10Rx2: 12 11 10 9Rx3: 12 11 10 9Rx4: 12 11 10Rx5: 12 11 10 9Rx6: 12 11 10 9Rx7: 12 11 10 9Rx8: 12 10

Error message

2247384 Diagnostic error. The signal's energy level is severely below the what is expected. See Long Description 2247384 tempmr diagCoord rrfRxFruIso.c 1539 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 18571 Object: APS Test Task Test: '16 Ch Switch Input Osc. AB, Port A' diagnostic error. The signal's energy level is severely below the what is expected. 'Dead' channels are: 1-8-> Extended Error: Sys Cab and UTNS Bypass Diag: Dead channels while running these diags indicate a problem with a board in the receive chain (MUX, UTNS, Rx), or cables between the boards. Sixteen bad channels (or eight bad channels in an eight channel system would indicate a problem with the signal source, or problems other than the three receive chain boards. Eight bad channels in a 16 channel system would indicate a bad board. Swapping good boards with suspected boards is an effective way to isolate a problem. Problems with banks of channels (1-4, 5-8, etc) may indicate a problem with a receive When individual bad channels (1,8,9,16) may lead to an improperly seated cable.\n 16 Channel Switch diag: If problems exist when running this diagnostic, the Sys Cab Loopback d

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5250094 RECEIVER 2A BOARD 5250094 RECEIVER 2A BOARD
Area
MR

Full cause & step-by-step resolution