GE Healthcare · MRExciteSigna224816322482172847174
RRF Calibration Error in RRF Data Path Diagnostics due to DRF board.
Symptom
RRF Calibration Error in RRF Data Path Diagnostics
Error message
2248163 Receiver calibration validation error. 2004 Error: 2248163 Host: OR3T Process: NSP File: APS:RrfCalibrator::rcvCal Line: 4036 Receiver calibration validation error. Significant measurement discrepancies exist between receivers. First detected: Rx1 vs. Rx2 2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2215142 DRF board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.