Pular para o conteúdo principal
GE Healthcare · MRExcite224821728471742248259

RRF Calibration Error Due to a Bad Connection From RRF MUX to RRF Exciter.

Symptom

Receiver Cal Errors logged on every System Boot Errors logged: Error: 2248217, 2248163. See Error Concepts for specifics.

Error message

2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10

2248259 VERY LOW SIGNAL STRENGTH IN RRF CAL. Low or no Signal Strength Seen From Exciter 0 (0=Master,1=Slave) Mon May 3 13:52:03 2004 Error: 2248259 Host: lx-mr Process: NSP File: APS:RrfCalibrator::excCal Line: 2541 VERY LOW SIGNAL STRENGTH IN RRF CAL Low or no signal strength seen from Exciter 0 (0=Master,1=Slave) during RRF cal.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2297729-3 Cable RRF Slot 7R-J6 RRF Master Exciter J6 to Mux Board J10
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues