Pular para o conteúdo principal
GE Healthcare · MRSigna22482172847174

Will not scan with error 2248217 Receiver calibration error.

Symptom

Will not scan with error 2248217 Receiver calibration error Fri Jul 18 05:59:36 2003 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APSReceiver calibration error(s). Receiver DB out of spec.R1 values that failed:Rx1: 12 10Rx2: 12 11 10 9Rx3: 12 11 10 9Rx4: 12 11 10Rx5: 12 11 10 9Rx6: 12 11 10 9Rx7: 12 11 10 9Rx8: 12 10

Error message

2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.