Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem Due to Loose Connection on AMP Data…
Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem Due to Loose Connection on AMP Data Bus Ribbon Cable.
Symptom
Will Not Scan with Gradient Errors.
Error message
2240243 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem See Long Description Fri Jan 28 07:30:43 2005 Error: 2240243 Host: lsdi Process: NSPFile: HPC_PORT_ID:spi_gca_error Line: 9 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect X amp bus cable and power on GS in local mode. If GS turns on, replace X Gradient Amp;else, check amp bus cable or SCA. 2240275 Z axis gradient amplifier fault 9 AMP NO.1 power module fault, over current or loss of clock, please check the clock cable between TYME2 and GPS 2240259 Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem Aug 3 Mon 08:20:27 2240259 Y Gradient Amplifier Fault.Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect Y amp bus cable and power on GS in local mode. If GS turns on, replace Y Gradient Amp; else, check amp bus cable or SCA.
Applies to
- Product
- Contour 3 & 4
- Software
- 7.6x Profile1 through 3 and Contour1 through 3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Intermittent SPI MDS LInk and X Gradient faults caused by the OC 5 vdc power supply output too low.
- Will Not Scan with Gradient Errors.
- Gradient Amplifier Fault.
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.