Pular para o conteúdo principal
GE Healthcare · MR223977522397742240243

Intermittent SPI MDS LInk and X Gradient faults caused by the OC 5 vdc power supply output too low.

Symptom

Scan Aborts with MDS Link Error

Error message

2239775 The SPI received an MDS broke error packet from the PACTL Error: 2239775 Host: oc001 Process: NSP File: HPC PORT ID:spi ctlx Line: 400 SPI MDS link failure. The SPI received an MDS broke error packet from the PACTL. The link appears to be broken between the GPS and the PACTL. Check GPS power, SCA(SCIF) board, & connections & signal level of fiber optic link between the GPS & CTLX.

2239774 SPI MDS link failure. The SPI received an MDS broke error packet from the GPS Sat Jan 29 08:33:06 2005 Error: 2239774 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 300 SPI MDS link failure The SPI received an MDS broke error packet from the GPS. The link appears to be broken between the SPI and the GPS. Check the connections and signal level of the MDS fiber optic link between the TYME card and the GPS.

2240243 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem See Long Description Fri Jan 28 07:30:43 2005 Error: 2240243 Host: lsdi Process: NSPFile: HPC_PORT_ID:spi_gca_error Line: 9 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect X amp 

Applies to

Product
Contour 3 & 4
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.