GE Healthcare · MR2249410236392227915
Unable to Prescan with RDBM, IT Link error Due To APS board with IT Daughterboard.
Symptom
Unable to Prescan with RDBM, IT Link error.
Error message
2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c 2227915 The RDBM Read Header function returned the following error: 6. The IT Link is possibly disconnected or the IT boards are not communicating.
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Component
- 2294300-4 APS BOARD WITH IT DAUGHTER BD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID102062. Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6.
- SCP board failes to send large packets
- Unable to Boot Applications Software with TPS Download Failure due to SCP2 Board in MGD Chassis.
- Data Acquisition errors during scan.
- One of the 2 MGD power supply cards had no leds lit.
- PSD download errors on any 8 channel coil Due to DRF Board.