Data Acquisition errors during scan.
Symptom
Data Acquisition errors during scan. Data acquisition error occurred. Run MGD datapath diagnostics, AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 000000ceH Also gradient hammer tests fail.
Error message
2253613 Datapath test detected unexpected data in sequence 0. Tue Jun 21 14:29:46 2005 Error: 2253613 Host: lx-mr Process: diagCoord File: srfToApDabSlave.c Line: 151 Diagnostic: SRF->AP DP Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 6730 Object: APS Test Task Datapath test detected unexpected data in sequence 0. APS local Address: 0 2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c
Applies to
- Models
- Excite
- Product
- Profile with Excite
- Software
- Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.
- Component
- 2274194 IRF Bd (Interface & Remote Functions Board) MGD2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.