Pular para o conteúdo principal
GE Healthcare · MRExcite2253613224941023639

Data Acquisition errors during scan.

Symptom

Data Acquisition errors during scan. Data acquisition error occurred. Run MGD datapath diagnostics, AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 000000ceH Also gradient hammer tests fail.

Error message

2253613 Datapath test detected unexpected data in sequence 0. Tue Jun 21 14:29:46 2005 Error: 2253613 Host: lx-mr Process: diagCoord File: srfToApDabSlave.c Line: 151 Diagnostic: SRF->AP DP Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 6730 Object: APS Test Task Datapath test detected unexpected data in sequence 0. APS local Address: 0

2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c

Applies to

Models
Excite
Product
Profile with Excite
Software
Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.
Component
2274194 IRF Bd (Interface & Remote Functions Board) MGD2
Area
MR