Pular para o conteúdo principal
GE Healthcare · MRExcite224941023639

One of the 2 MGD power supply cards had no leds lit.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2294300-8 MGD Power Supply ***OBSOLETE PART Case # 11430551*** See Long Description ***OBSOLETE PART Case # 11430551*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-254 Suitable alternative part is 5181495-3. Please see latest appropriate FRU manual for product in question. 212002950 February 20, 2009.
Area
MR

Full cause & step-by-step resolution