CWA ID102062. Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6.
Symptom
Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6 The IT link is possibly disconnected or the IT Boards are not communicating See Long Description The system gives the following errors immediately once autoprescan is passed and scan begins. This happens on all coils once this error occurs. 2227915 "The RDBM Read Header function returned the following error :6 The IT link is possibly disconnected or the IT Boards are not communicating." 2249422 "An ermes-triggered error was detected in task dir (0x396E558) for ermes number 2227915. Additional info in /usr/g/service/log/vxermesTrgLg-9102" Ran the IT diags and got the following error: 2252055 "IT-Host Card driver initialization error. This error usually means that the IT cable is disconnected or that the cable is damaged." PQR 13065908
Error message
2227915 The RDBM Read Header function returned the following error: 6. The IT Link is possibly disconnected or the IT boards are not communicating.
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.