GE Healthcare · MRSigna2218592224941060265224813022481222218630
PSD download errors on any 8 channel coil Due to DRF Board.
Symptom
Unable to Scan With The Download of a Prescribed Series Failed Message
Error message
2218592 The MrMail send of real time filter to TPS timed out Thu Jun 3 05:32:17 2004 2218592 qstmr1 scn msaUtil.m 175 The MrMail send of real time filter to TPS timed out. 2249410 AGP A PCI error occured. The MCP750 received a Target Abort Error: 2249410 Host: lx-mr Process: diagCoord File: excLog.c Line: 1151 Task ID: 0x3ac1d68 Time: 60265 Object: AGP A PCI error occured. The MCP750 received a Target Abort. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x1. Raven Err Address Reg: 0x1D0 2248130 APS DC Setup DRF Channels Error. DRFs can only support -x channels 2248130 Host: UMIMR3 Process: NSP File: APS:DacqControl::processPkt Line: 513 APS DC setup DRF channels error DRFs can only support -1 channels Software error, please perform a TPS Reset 2248122 Number of Receiver Prescribed is Over the DRF Capability. 2248122 Host: UMIMR3 Process: NSP File: APS:DrfDriver::setupChannel Line: 20 Number of receiver prescribed is over the DRF capability. Maximum receivers supported at this system is 7 Please change prescription. 2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed.
Applies to
- Models
- Signa
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.