Pular para o conteúdo principal
GE Healthcare · MRSigna2218592224941060265224813022481222218630

PSD download errors on any 8 channel coil Due to DRF Board.

Symptom

Unable to Scan With The Download of a Prescribed Series Failed Message

Error message

2218592 The MrMail send of real time filter to TPS timed out Thu Jun 3 05:32:17 2004 2218592 qstmr1 scn msaUtil.m 175 The MrMail send of real time filter to TPS timed out.

2249410 AGP A PCI error occured. The MCP750 received a Target Abort Error: 2249410 Host: lx-mr Process: diagCoord File: excLog.c Line: 1151 Task ID: 0x3ac1d68 Time: 60265 Object: AGP A PCI error occured. The MCP750 received a Target Abort. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x1. Raven Err Address Reg: 0x1D0

2248130 APS DC Setup DRF Channels Error. DRFs can only support -x channels 2248130 Host: UMIMR3 Process: NSP File: APS:DacqControl::processPkt Line: 513 APS DC setup DRF channels error DRFs can only support -1 channels Software error, please perform a TPS Reset

2248122 Number of Receiver Prescribed is Over the DRF Capability. 2248122 Host: UMIMR3 Process: NSP File: APS:DrfDriver::setupChannel Line: 20 Number of receiver prescribed is over the DRF capability. Maximum receivers supported at this system is 7 Please change prescription.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. 

Applies to

Models
Signa
Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.