Auto and Manual Prescan Fails After Timing Out.

Symptom

Auto and Manual Prescan Fails After Timing Out. The Scanner Does Not Pulse. TPS Reset Fails With Error Code 2226595 Unable to measure a valid line frequency See Long Description Auto or Manual Prescan fails after timing out. The scanner does not pulse. The two autoprescan errors shown below are logged. If a TPS reset is attempted, the last error is logged. These errors are logged when autoprescan is attempted: 2224583 MRSC_OC0 NSP TPS:aps_cf 15 Auto Prescan failed, possible data acquisition failure. 2224666 MRSC_OC0 psc PscMainControl.cc 1266 Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan. ============================================== This error is logged when TPS reset is attempted: 2226595 MRSC_OC0 NSP IPG:line_freq_check.c 184 IPG INIT: Unable to measure a valid line frequency; thus, cannot verify 'lineFreq' value in /w/config/MRconfig.cfg. Make sure this value is set correctly. Run LCA-related IPG diags.

Error message

2226595 Unable to measure a valid line frequency See Long Description 2226595 MRSC_OC0 NSP IPG:line_freq_check.c 184 IPG INIT: Unable to measure a valid line frequency; thus, cannot verify 'lineFreq' value in /w/config/MRconfig.cfg. Make sure this value is set correctly. Run LCA-related IPG diags.

2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.

2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure.

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.