GE Healthcare · MR223977522256932239774
MDS Link faults due to no power to GPS or RF Amp in the IPS.
Symptom
TPS or MGD Reset Fails With MDS Link Errors
Error message
2239775 The SPI received an MDS broke error packet from the PACTL Error: 2239775 Host: oc001 Process: NSP File: HPC PORT ID:spi ctlx Line: 400 SPI MDS link failure. The SPI received an MDS broke error packet from the PACTL. The link appears to be broken between the GPS and the PACTL. Check GPS power, SCA(SCIF) board, & connections & signal level of fiber optic link between the GPS & CTLX. 2225693 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B 2225693 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B 2239774 SPI MDS link failure. The SPI received an MDS broke error packet from the GPS Sat Jan 29 08:33:06 2005 Error: 2239774 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_scif Line: 300 SPI MDS link failure The SPI received an MDS broke error packet from the GPS. The link appea
Applies to
- Product
- Profile 4
- Software
- 7.8x Profile4 or ProfileHSR only
- Component
- 2112116 PDU Relay Assy For Profile
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Accidental Push of E-stop Button.
- CWA ID115238. Scan Aborts with MDS Link Error.
- MDS Link faults and Loss of AC Mains.
- Intermittent SPI MDS LInk and X Gradient faults caused by the OC 5 vdc power supply output too low.
- Unstable Buck Convert / H Bridge Power Supply in SSM.
- CWA Gradient Amplifier Fails to go to Ready.