CWA Gradient Amplifier Fails to go to Ready.

Symptom

Gradient Amplifier Fails to go to Ready

Error message

2225336 The SPI received a packet indicating that SRI has reset. See Long Description Sun May 24 15:28:19 2009 2225336 MR4 NSP HPC_PORT_ID:spi_sri_recover_link 26 SRI-SPI Communication Error. The SPI received a packet indicating that SRI has reset.

2238489 The MIF is reporting a long overload on the Y axis HIGH 8645 Power

2224666 Auto Prescan failed, Gradient amplifiers are not ready

2225693 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B 2225693 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR