GE Healthcare · MR2240311
Unable to scan due to error code 2240311 FLT AC PHASE.
Symptom
Unable to scan due to error code 2240311 FLT AC PHASE See long description. Lx 9.1 Base Horizon system with 8280 gradient amps logging error code 2240311 "FLT AC PHASE" after room got too hot when air conditioning was off. The room temperature is normal now but problem remains. This message also appears on the SCA display. Input phases are okay. FE checked and reseated the PSU Bus cable, but problem remains. These amps are also used on HFO systems.
Error message
2240311 FLT AC PHASE Does not apply to Signa Profile This message also appears on the SCA display. 2240311 FLT AC PHASE
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- MDS Link faults and Loss of AC Mains.
- CWA 2240311 Gradient Power Supply Unit Fault due to tripped breaker on the Gradient Power Supply.
- Unable to scan due to error code 2240311 "FLT AC PHASE"
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.