Malfunction of control system data communication — MUDAT ROTSOT/SOTROT fault tree (Data Transmission)
Use this page when the console displays "A data transmission section error occurred." This fault tree walks from that message to a specific abnormal PWB (OPCONTA, GCIFA, RPDM, SOT, or ROT) or a cable-connection problem, using the built-in MUDAT ROTSOT/SOTROT tests and the two PWB loopback tests (OPCONTA test, GCIFA test) already defined in the data-transfer test-tool reference (Related).
Applies to
Activion16 (TSX-031A). Volume 2, Data Transmission Unit chapter, §3.5.2. Source does not call out a hardware-generation-specific variant of this fault tree — it applies to the OPCONTA/GCIFA/MUDAT data-transfer path as described throughout the chapter.
Steps
14 steps
Source
2D201-113EN*J §3.5.2 'Malfunction of control system data communication' (pp.362-366)
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| OPCONTA (rotation-side interface PWB) | n/a in source | |
| GCIFA (stationary-side interface PWB) | n/a in source | |
| RPDM PWB | n/a in source | |
| SOT PWB (LPBM) | n/a in source | |
| ROT PWB (LPBA) | n/a in source | |
| PRTM | n/a in source | |
| PPBM | n/a in source | |
| SPD0 | n/a in source | |
| KGTSM PWB | n/a in source |
Full step-by-step for Malfunction of control system data communication — MUDAT ROTSOT/SOTROT fault tree (Data Transmission)
All 14 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Data Transmission guides
Check gantry rotation and MUDAT data transfer after working on the MUDAT7 stepsCheck the MUDAT transmission-LED luminescence (ROT and SOT sides)7 stepsData transfer section error-identification test tools (DAS / Console / ROTSOT / SOTROT / OPCONTA / GCIFA)6 stepsDisassemble the MUDAT (rotation and stationary section PWBs)8 stepsGCIFA-D2 PWB troubleshooting flowchart (Data Transmission)0 stepsMalfunction of DAS data communication — cause investigation (streak/ring artifacts)26 steps
View all Activion16 (TSX-031A) guidesSource: 2D201-113EN*J §3.5.2 'Malfunction of control system data communication' (pp.362-366). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.