Pular para o conteúdo principal
ProcedureData TransmissionFunctional ChecksCheck

Check the MUDAT transmission-LED luminescence (ROT and SOT sides)

The light from the MUDAT's communication LEDs is infrared — invisible to the eye. To check whether a given LED is actually emitting, convert its infrared light to visible red light with an IR converter and look for the red glow. This page covers both halves of that check: the ROT side (rotation-section transmission LEDs) and the SOT side (stationary-section transmission LEDs).

Applies to
Activion16 (TSX-031A), Data Transmission Unit — MUDAT rotation-section (ROT) and stationary-section (SOT) transmission LEDs, accessed via the connector cover at the rear of the SMUDAT. No model- or revision-specific restriction is stated in source for this check.
Steps
7 steps
Source
2D201-113EN*J §3.5.4 'Procedure for checking the luminescence' (pp.391-394)

Safety

DANGER

Hazard: Hazardous voltage in the LCSR (slipring section) and brush section, behind the rear dome cover

Consequence: Electric shock

Avoidance: Do not remove the rear dome cover. If it is necessary to remove the rear dome cover, turn OFF the breaker before mounting/removing it. Do not turn ON the power of the gantry when the rear dome cover is removed.

WARNING

Hazard: Inadvertent gantry rotation during servicing work while checking the ROT-side or SOT-side LEDs

Consequence: Source states this precaution exists "to ensure safety of the workers," implying injury risk to personnel near the rotation section

Avoidance: Use the cover switch, CP320, rotation mechanical stopper, etc., to prevent inadvertent gantry rotation during servicing work (printed identically in the ROT-side and the SOT-side WARNING). A separate SOT-side body instruction additionally cites CP320 and a mechanical rotation lock pin, etc. — rendered as its own step, not as a WARNING variant.

NOTICE

Hazard: Rotating the rotation section manually without releasing the brake first (ROT-side closeout)

Consequence: May result in system malfunction (source states an equipment-only consequence; printed under a CAUTION heading, re-leveled to NOTICE per Z535)

Avoidance: Always release the brake before rotating the rotation section manually

NOTICE

Hazard: Replacing or checking static-sensitive parts (PWBs, hard disks, etc.) without ESD protection (SOT-side closeout)

Consequence: Damage to static-sensitive components (source prints this under a CAUTION heading; re-leveled to NOTICE per Z535 because the only stated consequence is to the parts, not to personnel)

Avoidance: Wear an antistatic wristband when replacing or checking parts that can be damaged by static electricity

Full step-by-step for Check the MUDAT transmission-LED luminescence (ROT and SOT sides)

All 7 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Data Transmission guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §3.5.4 'Procedure for checking the luminescence' (pp.391-394). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.