ProcedureData TransmissionSetup & CalibrationSetup

Set the GCIFA tube-position switches for a Disable (non-rotation) scan

Configure the GCIFA position switches before running a Disable (non-rotation) scan from the DCA. Because gantry rotation is not performed during a Disable scan, the MUDAT position data that the OPCONTA would normally receive from the KGTSM (via the GCIFA, in step with gantry rotation — see the IF-unit operating principle, Related) simply does not exist. This procedure substitutes a manual position

Applies to
Activion16 (TSX-031A) data transfer section (OPCONTA/GCIFA/GCIFA-D2, MUDAT link). Used only when running a Disable (non-rotation) scan via the DCA — i.e., when the gantry is deliberately not rotating and the normal KGTSM-generated MUDAT position data (§3.3.2(4)) cannot be obtained. No further model- or revision-specific restriction is stated in source for this subsection.
Steps
4 steps
Source
2D201-113EN*J §3.3.2(5) 'Non-rotation (Disable) scan' (p.333)

Full step-by-step for Set the GCIFA tube-position switches for a Disable (non-rotation) scan

All 4 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Data Transmission guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §3.3.2(5) 'Non-rotation (Disable) scan' (p.333). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.