TroubleshootingSoftwareTroubleshooting
Test Image STI2
Print the STI2 test image and check its 3 fields at the lightbox — a fine line/dot pattern, an embedded SMPTE reference pattern, and a Dmax 1.0 flatfield — to catch most image-quality faults the DryStar 5300 can produce in one printout.
Applies to
DryStar 5300 (Type 5365/100)
Personnel
1 technician · Service-level technician — navigates the Service Engineer Tools > Enter Service Command Session webpages, prints the STI2.ZIP test image, and evaluates the 3 resulting fields at the lightbox; no disassembly required for this entry itself (derived — source does not assign a formal skill-level label to this section).
Steps
2 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service: Machine specific tools, software tools and auxiliary means, Chapter 3.2/13, §2.3.2 'Testimage STI2', Edition 1, Revision 7 (2006-01-26), p.129
Full step-by-step for Test Image STI2
All 2 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Software guides
ADVP — DICOM Test Tool6 stepsPrinter Bootup-Message via HyperTerminal12 stepsEmergency SW Installation on a New, Empty Internal CF Card6 stepsEmergency SW Installation on a Printer in "Error - System Failure" State4 stepsEntering Site Data0 stepsAdjusting Image Quality at the Printer (LUT, Interpolation, Dmax)6 steps
View all DryStar 5300 guidesSource: DD+DIS346.05E — Drystar 5300 Repair and Service: Machine specific tools, software tools and auxiliary means, Chapter 3.2/13, §2.3.2 'Testimage STI2', Edition 1, Revision 7 (2006-01-26), p.129. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.