TH Driver Banding, Drop & Intermittent Drop
Three visually different line faults in the STI2 test image — a persistent dark line, a persistent blank line, and an intermittent blank line — all trace back to the same cause and the same fix: a defective thermal-head (TH) driver, corrected by replacing the Thermal Head.
Applies to
DryStar 5300 (Type 5365/100)
Personnel
1 technician · Basic — visual pattern match against the STI2 test image, then a label/infocounter check to identify the TH type before ordering the replacement part; the TH replacement itself is a separate procedure.
Steps
3 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service, Troubleshooting, Chapter 3.3 (item 11.2.3, cont'd — TH driver sub-symptoms), p.185-187, Edition 1 Rev.7
Full step-by-step for TH Driver Banding, Drop & Intermittent Drop