Skip to main content
Alarm — stator control board error conditionRotor Control Board

ER32 — Rotor Control Board ARM processor error

Philips CT error code, reported by the Rotor Control Board.

What ER32 means

On Access Series CT (6/16/Dual) CT systems, status code ER32 indicates: Rotor Control Board ARM processor error. The code originates in the Rotor Control Board section of the Rotor Control Board. Severity level alarm (Alarm — stator control board error condition).

Reported action: Check the Rotor Control Board LED indicators (Table 1-1) for the ARM/Spartan communication LEDs (V12, V18); if the fault persists, replace the Rotor Control Board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ER32

The complete diagnosis, cause tree, and step-by-step remedy for the Rotor Control Board subsystem are documented in the FSELIB Philips service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related Rotor Control Board codes

Discuss ER32 with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.