200011E3Scanhead configuration problem

Philips CT error code, reported by the Scanhead / probe.

What 200011E3 means

On CT systems, status code 200011E3 indicates: Scanhead configuration problem. The code originates in the Scanhead / probe section of the Scanhead / probe. Severity level warning (Warning — scanhead configuration).

Reported action: Confirm the connected scanhead is one the system is configured to support and that the selected transducer matches the port in use. Reseat the scanhead connector and check its pins, then reselect the probe. If only one scanhead or one port produces the code, that isolates the fault to that probe or port..

From the Philips ultrasound service reference "Error Logger Overview" (New Error Logger), archived under Philips ultrasound End-of-Life service documentation. The source lists this Errno as "Scanhead-related configuration problem". A full log entry combines a 4-digit Subsystem code with the Errno (format ssss-eeeeeeee).