What FD030000 means

On Aquilion RX / V4-62R004 and Aquilion ONE / V4.2 CT systems, status code FD030000 indicates: Error in recon-process — RTM rRecCtl (Genesis). The code originates in the RTM — rRecCtl for Genesis section of the RTM (real time manager) — rRecCtl for Genesis. Severity level A0 (Alarm).

Reported action: Note the fault time and the study whose reconstruction failed, then retry the reconstruction from the stored raw data and read the RTM error log entries recorded with the alarm. If reconstruction keeps failing, follow the RTM troubleshooting procedure in the service manual..

Toshiba/Canon Aquilion RX (V4-62R004) and Aquilion ONE (V4.2) CT systems. The service error-message table lists this code under the "RTM rRecCtl for Genesis" device group (set 204), with the OEM text "Error in recon-process." The same message text also appears under codes FD010000, FD020000, FD040000, FD050000 and FD060000 in that same device group.

Full troubleshooting for FD030000

The complete diagnosis, cause tree, and step-by-step remedy for the RTM — rRecCtl for Genesis subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM — rRecCtl for Genesis codes

Related service procedures

Discuss FD030000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, Aquilion RX V4-62R004 / Aquilion ONE V4.2). Reference information re-authored by FSELIB — not a copy of the OEM manual.