Pular para o conteúdo principal
AlarmRTM — rRecCtl

C0830001Reconstruction process error reported by the RTM (rRecCtl)

Canon Medical (Toshiba) CT error code, reported by the RTM (Real Time Manager) — rRecCtl.

What C0830001 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0830001 indicates: Reconstruction process error reported by the RTM (rRecCtl). The code originates in the RTM — rRecCtl section of the RTM (Real Time Manager) — rRecCtl. Severity level A0 (Alarm).

Reported action: Note the time of the alarm and check the reconstruction job and its log for the affected scan, then retry the reconstruction. If reconstruction keeps failing, check the RTM board for related alarms and follow the service manual's RTM troubleshooting guidance..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems. The service error-message table lists this code under the "Device RTM rRecCtl" group with the message "Error in recon-process." The same message text is also listed for codes C0830002, C0830003, C0830004, C0830005 and C0830006. The RTM (real time manager) is the console's master real-time control board; on the Aquilion 16 it sits in the cPCI chassis and controls the cPCI chassis, the SS/ADI and the gantry.

Full troubleshooting for C0830001

The complete diagnosis, cause tree, and step-by-step remedy for the RTM — rRecCtl subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM — rRecCtl codes

Related service procedures

Discuss C0830001 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng), device group "Device RTM rRecCtl". Reference information re-authored by FSELIB — not a copy of the OEM manual.