Pular para o conteúdo principal
AlarmRTM PRE — pre-process stage

C0010015Standard pre-process alarm on the CT console (RTM PRE)

Canon Medical (Toshiba) CT error code, reported by the RTM PRE (pre-process stage).

What C0010015 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code C0010015 indicates: Standard pre-process alarm on the CT console (RTM PRE). The code originates in the RTM PRE — pre-process stage section of the RTM PRE (pre-process stage). Severity level A0 (Alarm).

Reported action: Look at the images produced by the affected exam and record the protocol that raised the alarm. The grading is a standard alarm (A0) and the table carries no code-specific detail beyond the pre-process error, so log a single occurrence and call your service representative if it starts repeating..

Reported on Aquilion RX/V4-62R004, Aquilion ONE/V4.2 and V3.20R003. It comes from the 100:Device RTM PRE group of the console error table, where the OEM text reads "Error in pre-process.", graded A0 — a standard alarm rather than the A1 high-severity grading carried by most of the sibling RTM PRE codes in the same table.

Full troubleshooting for C0010015

The complete diagnosis, cause tree, and step-by-step remedy for the RTM PRE — pre-process stage subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM PRE — pre-process stage codes

Related service procedures

Discuss C0010015 with other field engineers in the FSELIB forum

Source: errmsg.eng — 100:Device RTM PRE table. Reference information re-authored by FSELIB — not a copy of the OEM manual.