A6060000 — Recon-process error reported by the RTM (Genesis)
Canon Medical (Toshiba) CT error code, reported by the RTM (real time manager) — REC (recon-process) for Genesis.
What A6060000 means
On Aquilion RX / V4-62R004 and Aquilion ONE / V4.2 CT systems, status code A6060000 indicates: Recon-process error reported by the RTM (Genesis). The code originates in the RTM — REC (recon-process) for Genesis section of the RTM (real time manager) — REC (recon-process) for Genesis. Severity level A1 (Alarm — high severity).
Reported action: Retry the reconstruction after noting the time stamp, then compare the recon-process log entries around that moment. Persistent faults should be diagnosed with the RTM troubleshooting procedure in the service manual..
Toshiba/Canon Aquilion RX (V4-62R004) and Aquilion ONE (V4.2) CT systems. The service error-message table lists this code in its "Device RTM REC for Genesis" device group, with the message text "Error in recon-process." The group covers the REC (recon-process) stage handled by the RTM (real time manager). The same message text is also listed in that group under codes A6030000, A6040000, A6050000, A6070000, A6080000, A6090000, A60A0000, A60B0000, A60C0000, A60D0000, A60E0000 and A60F0000.
Full troubleshooting for A6060000
The complete diagnosis, cause tree, and step-by-step remedy for the RTM — REC (recon-process) for Genesis subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated RTM — REC (recon-process) for Genesis codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng), device group "Device RTM REC for Genesis". Reference information re-authored by FSELIB — not a copy of the OEM manual.