AlarmPower Supply Module — Over-Temperature Protection

HC11329Power supply module over-temperature protection tripped

Agfa CT error code, reported by the Power Supply Module.

What HC11329 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11329 indicates: Power supply module over-temperature protection tripped. The code originates in the Power Supply Module — Over-Temperature Protection section of the Power Supply Module. Severity level alarm (Alarm).

Reported action: Check cooling conditions first — dust buildup or a failed cooling fan — and let the module cool. If the warning keeps recurring after cooling is confirmed adequate, treat the power supply module as defective: power off and replace it (Part PS_spare)..

Raised on the Drystar 5500/5503 when the power supply module's own over-temperature protection trips. Classed as a Failure — a warning is issued but no immediate service intervention is required. Local display shows "PWS-overtemperature. Cooling down.", distinct from the Defect-level HC10301, which uses near-identical Meaning wording ("Temperature of powersupplymodule is too high") but points straight to "Call service".

Full troubleshooting for HC11329

The complete diagnosis, cause tree, and step-by-step remedy for the Power Supply Module — Over-Temperature Protection subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Power Supply Module — Over-Temperature Protection codes

Discuss HC11329 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.