HC11238 — Sorter entry sensor S25 stuck activated at startup (film jam)
Agfa CT error code, reported by the Output Sorter Module.
What HC11238 means
On DryStar 5500 and DryStar 5503 CT systems, status code HC11238 indicates: Sorter entry sensor S25 stuck activated at startup (film jam). The code originates in the Output Sorter Module — Jam / Timing Faults section of the Output Sorter Module. Severity level alarm (Alarm).
Reported action: Check the sorter section and clear the film jam at sensor S25. If this keeps recurring, look for a mechanical blockage in the sheet transport path, or a broken/misfit sensor..
Raised on the DryStar 5500/5503 output sorter module when the sorter entry sensor S25 is activated continuously during printer startup/initialization or during the empty-check procedure, because a film remained at that sensor location. Filed with the other output-sorter fault codes (HC10823, HC10831-HC10835, HC10907-HC10909, HC11223-HC11224, HC11238-HC11241).
Full troubleshooting for HC11238
The complete diagnosis, cause tree, and step-by-step remedy for the Output Sorter Module — Jam / Timing Faults subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Output Sorter Module — Jam / Timing Faults codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.