Pular para o conteúdo principal
AlarmFilm Tray Pick-up (Lower) — Passed Sensor S20

HC11230Lower pick-up passed sensor S20 stuck on (continuous activation) at startup

Agfa CT error code, reported by the Film Tray Pick-up Module (Lower Tray).

What HC11230 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11230 indicates: Lower pick-up passed sensor S20 stuck on (continuous activation) at startup. The code originates in the Film Tray Pick-up (Lower) — Passed Sensor S20 section of the Film Tray Pick-up Module (Lower Tray). Severity level alarm (Alarm).

Reported action: Check the lower input tray and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..

Raised on the Drystar 5500/5503 lower film tray pick-up module when the lower pick-up passed sensor S20 stays continuously activated during startup/initialization or the empty-check procedure. Its time-out variant is HC11204; the upper-tray equivalent is HC11229.

Full troubleshooting for HC11230

The complete diagnosis, cause tree, and step-by-step remedy for the Film Tray Pick-up (Lower) — Passed Sensor S20 subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Film Tray Pick-up (Lower) — Passed Sensor S20 codes

Discuss HC11230 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.