HC11227 — Ambient temperature too low, measured on the MEC board (image quality only)
Agfa CT error code, reported by the Print Drum Module.
What HC11227 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11227 indicates: Ambient temperature too low, measured on the MEC board (image quality only). The code originates in the MEC Board — Ambient Temperature Sensor section of the Print Drum Module. Severity level alarm (Warning).
Reported action: If the printer was recently in a cold environment, allow it to warm up before troubleshooting further; if the site's ambient temperature is genuinely too low, it must be raised above the threshold before the condition clears. No hardware fix is documented for this code..
Raised on the Drystar 5500/5503 when the ambient-temperature reading taken on the MEC board is too low. Classed as Accidental — the engine keeps printing, but image quality may decrease. Local display shows "Amb. temp. low". Distinct from HC10817, which measures ambient temperature on the THIB and does stop printing.