HC11225 — Homing timeout during initialization
Agfa CT error code, reported by the System Initialization.
What HC11225 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11225 indicates: Homing timeout during initialization. The code originates in the System Initialization — Homing Timeout section of the System Initialization. Severity level alarm (Alarm).
Reported action: Check first around the drum-positioning motor, its gear and belt; then check the full film path for any film left jammed in the engine. If the timeout recurs after both checks come back clean, the fault is more likely in one of the other three homing groups (alignment module, pick-up magazine, or sorter positioning)..
Raised on the Drystar 5500/5503 during initialization when a homing sequence times out. The source names four possible origins for this single timeout: the horizontal/vertical alignment module, the pick-up magazine, the drum module, or the sorter positioning — but its own documented Cure points first at just one of them. Classed as Accidental.