Pular para o conteúdo principal
AlarmMDM (Macrodensitometer) Module — Takeover Mechanism

HC11221Takeover module fails to close

Agfa CT error code, reported by the MDM (Macrodensitometer) Module.

What HC11221 means

On DryStar 5500 and DryStar 5503 CT systems, status code HC11221 indicates: Takeover module fails to close. The code originates in the MDM (Macrodensitometer) Module — Takeover Mechanism section of the MDM (Macrodensitometer) Module. Severity level alarm (Alarm).

Reported action: Check the Takeover module and motor M9 for a mechanical obstruction preventing it from closing, and review the related fault that triggered this detection, since the Takeover module has no dedicated close sensor of its own..

Raised on the DryStar 5500/5503 when the Takeover module (part of the MDM sheet path, driven by motor M9) fails to close. The Takeover module has no dedicated sensors, so this detection is linked to another malfunction. Filed with the other MDM/densitometer fault codes (HC10200, HC10819-HC10820, HC10904-HC10905, HC10920, HC11005, HC11110-HC11115, HC11124, HC11218-HC11222).

Full troubleshooting for HC11221

The complete diagnosis, cause tree, and step-by-step remedy for the MDM (Macrodensitometer) Module — Takeover Mechanism subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MDM (Macrodensitometer) Module — Takeover Mechanism codes

Discuss HC11221 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.