WarningTH Module — Thermistor Temperature Measurement

HC11119Thermal printhead heatsink temperature too low

Agfa CT error code, reported by the Thermal Print Head (TH) Module.

What HC11119 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11119 indicates: Thermal printhead heatsink temperature too low. The code originates in the TH Module — Thermistor Temperature Measurement section of the Thermal Print Head (TH) Module. Severity level warning (Warning).

Reported action: Make sure the ambient temperature is within the printer's specifications, and let the printer acclimatize if it has just come from a cold environment. If the reading is still low after that, the cause is either one or more defective thermistors on the thermal head or an incorrect heatsink-temperature measurement on the MEC board — power off the printer and replace the thermal head (Part TH) or the MEC board, whichever the follow-up check points to..

Raised on the Drystar 5500/5503 thermal print head (TH) module when the heatsink-temperature measurement of the thermal printhead is too low. The engine is still allowed to print but image quality may decrease. Paired in the appendix with the equivalent high-temperature code, HC11120.

Full troubleshooting for HC11119

The complete diagnosis, cause tree, and step-by-step remedy for the TH Module — Thermistor Temperature Measurement subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related TH Module — Thermistor Temperature Measurement codes

Discuss HC11119 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.