HC11114 — MDM scan head fails to leave the home position within the time-out
Agfa CT error code, reported by the MDM (Macrodensitometer) Module.
What HC11114 means
On DryStar 5500 and DryStar 5503 CT systems, status code HC11114 indicates: MDM scan head fails to leave the home position within the time-out. The code originates in the MDM (Macrodensitometer) Module — Home Position section of the MDM (Macrodensitometer) Module. Severity level alarm (Alarm).
Reported action: Check for a physical jam that is keeping the MDM scan head from leaving its home position and remove the obstruction. If the scan head still will not move, the MDM scan head transport mechanism may be broken (replace the MDM assembly) or the home sensor S7 may be defective (replace S7)..
Raised on the DryStar 5500/5503 when the MDM (macrodensitometer) scan head is not detected leaving its home position within the time-out period. The printer shows "Calibration module not working". Filed with the other MDM/densitometer fault codes (HC10200, HC10819-HC10820, HC10904-HC10905, HC10920, HC11005, HC11110-HC11115, HC11124, HC11218-HC11222).