Pular para o conteúdo principal
AlarmMDM (Macrodensitometer) Module — Home Position

HC11113MDM scan head fails to return to the home position after a scan

Agfa CT error code, reported by the MDM (Macrodensitometer) Module.

What HC11113 means

On DryStar 5500 and DryStar 5503 CT systems, status code HC11113 indicates: MDM scan head fails to return to the home position after a scan. The code originates in the MDM (Macrodensitometer) Module — Home Position section of the MDM (Macrodensitometer) Module. Severity level alarm (Alarm).

Reported action: Check for a physical jam that is keeping the MDM scan head from returning home — if found, remove the obstruction and power-cycle the printer so the scan head re-homes. If there is no obstruction, the MDM scan head transport mechanism may be broken (replace the MDM assembly) or the home sensor S7 may be defective (replace S7)..

Raised on the DryStar 5500/5503 when the MDM (macrodensitometer) scan head is not detected returning to its home position after performing a scan. The printer shows "Calibration module not working". Filed with the other MDM/densitometer fault codes (HC10200, HC10819-HC10820, HC10904-HC10905, HC10920, HC11005, HC11110-HC11115, HC11124, HC11218-HC11222).

Full troubleshooting for HC11113

The complete diagnosis, cause tree, and step-by-step remedy for the MDM (Macrodensitometer) Module — Home Position subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MDM (Macrodensitometer) Module — Home Position codes

Discuss HC11113 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.