Pular para o conteúdo principal
AlarmAlignment Module — Film-Format Solenoid (EM2)

HC1091910-inch film-format alignment solenoid EM2 not connected or malfunctioning

Agfa CT error code, reported by the Alignment Module.

What HC10919 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10919 indicates: 10-inch film-format alignment solenoid EM2 not connected or malfunctioning. The code originates in the Alignment Module — Film-Format Solenoid (EM2) section of the Alignment Module. Severity level alarm (Alarm).

Reported action: Check the connector of solenoid EM2, then connector P27 on the MEC board — reconnect or replace whichever isn't OK. If the connections are fine, check the wiring from EM2 to MEC/P27 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace EM2; if the fault persists, the MEC board is defective and needs replacing..

Raised on the Drystar 5500/5503 when the alignment module's 10-inch film-format align solenoid EM2 is not connected or is malfunctioning. The printer shows "Call service".

Full troubleshooting for HC10919

The complete diagnosis, cause tree, and step-by-step remedy for the Alignment Module — Film-Format Solenoid (EM2) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Alignment Module — Film-Format Solenoid (EM2) codes

Discuss HC10919 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.