HC10905MDM scan motor M11 not connected or malfunctioning

Agfa CT error code, reported by the MDM / Densitometer Module.

What HC10905 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10905 indicates: MDM scan motor M11 not connected or malfunctioning. The code originates in the MDM Module — Scan Motor (M11) section of the MDM / Densitometer Module. Severity level alarm (Alarm).

Reported action: Check connector M11 and reconnect or replace if faulty; check connector P28 on the MEC board and reconnect or replace if faulty; check the wiring from M11 to MEC/P28 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace the MDM assembly; if the fault remains, the motor function on the MEC board has failed and the MEC board has to be replaced..

Raised on the Drystar 5500/5503 MDM (macrodensitometer) module when the scan motor M11 is not connected or is malfunctioning. The printer shows "Calibration module not working".

Full troubleshooting for HC10905

The complete diagnosis, cause tree, and step-by-step remedy for the MDM Module — Scan Motor (M11) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.